Summary of Source and Measure Functions | |||||
Function | Measure | Source | |||
DC V | 0~30V | Not available | |||
Resistance | 0~3200Ω | 0~3200Ω | |||
RTD | Pt100,Pt1000,Cu50,Cu100 | Pt100,Pt1000,Cu50,Cu100 | |||
Others | Stepping output, ramp, phase step output, user-defined range | ||||
1,DC Voltage Measurement | |||||
Range | Maximum measurement range | Resolution | Accuracy (%of reading+Count) | ||
30V | 0V~31V | 0.001V | 0.05+2 | ||
2,Resistance Measurement | |||||
Range | Maximum measurement range | Resolution | Accuracy(Ω) | ||
2-wire, 3-wire | 4-wire | ||||
400Ω | 0~440Ω | 0.01Ω | 0.25 | 0.15 | |
3200Ω | 420Ω~3600Ω | 0.1Ω | 1.5 | 1 | |
3,Resistance Output | |||||
Range | Maximum output range | Resolution | External exciting current | Accuracy (Ω) | |
400Ω | 0~440Ω | 0.01Ω | 0.4mA~4.0mA | 0.25 | |
3200Ω | 400~3600Ω | 0.1Ω | 0.1mA~0.5mA | 1 | |
4,RTD | |||||
Graduation | Range | Resolution | Accuracy(℃) | ||
2-wire,3-wire | 4-wire | Output | |||
Pt100 | -200℃~850℃ | 0.1℃ | 0.7 | 0.4 | 0.7 |
Pt1000 | -200℃~650℃ | 0.4 | 0.3 | 0.3 | |
Cu50 | -50℃~150℃ | 1.2 | 0.8 | 0.8 | |
Cu100 | -50℃~150℃ | 0.7 | 0.4 | 0.4 |