Summary of Source and Measure Functions | ||||
Function | Measure | Source | ||
DC V | 0~30V | Not available | ||
DC mV | 0~120mV(the upper display ± 24mA) | 0mV~100mV | ||
Resistance | 0~3200Ω | 0~3200Ω | ||
RTD | Pt100,Pt1000,Cu50,Cu100 | Pt100,Pt1000,Cu50,Cu100 | ||
TC | E,J,K,T,B,R,S,N | E,J,K,T,B,R,S,N | ||
Others | Stepping output,ramp, phase step output, user-defined range | |||
DC Voltage Measurement | ||||
Range | Maximum measurement range | Resolution | Accuracy (%of reading+counts) | |
30V | 0V~31V | 0.001V | 0.02+2 | |
100mV(lower display) | -15mV~125mV | 0.01mV | 0.02+2 | |
Resistance measurement | ||||
Range | Maximum measurement range | Resolution | Accuracy(Ω) | |
Two-wire, three-wire | Four-wire | |||
400Ω | 0~440Ω | 0.01Ω | 0.15 | 0.1 |
3200Ω | 420Ω~3600Ω | 0.1Ω | 1 | 0.5 |
Temperature, TC | ||||
Graduation | Range | Resolution | Accuracy(Ω) | |
J | -200℃~0℃ | 0.1℃ | 1.0℃ | |
0℃~1200℃ | 0.7℃ | |||
K | -200℃~0℃ | 0.1℃ | 1.2℃ | |
0℃~1370℃ | 0.8℃ | |||
T | -200℃~0℃ | 0.1℃ | 1.2℃ | |
0℃~400℃ | 0.8℃ | |||
E | -200℃~0℃ | 0.1℃ | 0.9℃ | |
0℃~950℃ | 0.7℃ | |||
R | -20℃~0℃ | 1℃ | 2.5℃ | |
0℃~500℃ | 1.8℃ | |||
500℃~1750℃ | 1.4℃ | |||
S | -20℃~0℃ | 1℃ | 2.5℃ | |
0℃~500℃ | 1.8℃ | |||
500℃~1750℃ | 1.5℃ | |||
B | 600℃~800℃ | 1℃ | 2.2℃ | |
800℃~1000℃ | 1.8℃ | |||
1000℃~1800℃ | 1.4℃ | |||
N | -200℃~0℃ | 0.1℃ | 1.5℃ | |
0℃~1300℃ | 0.9℃ |