Summary of Source and Measure Functions | |||
Function | Measure | Source | |
DC V | 0~30V | Not available | |
DC mV | 0~100mV | 0~100mV | |
Thermocouple | E,J,K,T,B,R,S,N | E,J,K,T,B,R,S,N | |
Others | Stepping output, ramp, phase step output,user-defined range | ||
1,DC Voltage Measurement | |||
Range | Maximum measurement range | Resolution | Accuracy (%of reading+Counts) |
30V | 0V~31V | 0.001V | 0.02+2 |
100mV | -15mV~125mV | 0.01mV | 0.02+2 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. | |||
Input resistance:>1MΩ. | |||
2,Thermocouple | |||
Graduation | Range | Resolution | Accuracy(Ω) |
J | -200°C~0°C | 0.1°C | 1.0°C |
0°C~1200°C | 0.7°C | ||
K | -200°C~0°C | 0.1°C | 1.2°C |
0°C~1370°C | 0.8°C | ||
T | -200°C~0°C | 0.1°C | 1.2°C |
0°C~400°C | 0.8°C | ||
E | -200°C~0°C | 0.1°C | 0.9°C |
0°C~950°C | 0.7°C | ||
R | -20°C~0°C | 1°C | 2.5°C |
0°C~500°C | 1.8°C | ||
500°C~1750°C | 1.4°C | ||
S | -20°C~0°C | 1°C | 2.5°C |
0°C~500°C | 1.8°C | ||
500°C~1750°C | 1.5°C | ||
B | 600°C~800°C | 1°C | 2.2°C |
800°C~1000°C | 1.8°C | ||
1000°C~1800°C | 1.4°C | ||
N | -200°C~0°C | 0.1°C | 1.5°C |
0°C~1300°C | 0.9°C | ||
Errors of cold-junction compensation are not included in the table.Accuracy of cold-junction compensation:1.5°C |